New TopShim run:

version = topshim 2.0.0
date = 2023-10-30
time = 14:31:17.497
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.3

initial B0 stdDev = 3.28 Hz
sample size = 2.87 cm, position = 0.01 cm
final B0 stdDev = 0.40 Hz > improvement = 8.3
envelope width = 0.66 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+61
Z2	-43
Z3	+124
Z4	-307
Z5	+59
X	-34
Y	+134

duration = 3 min 37 sec
completed successfully
finished Mon Oct 30 14:34:53 2023


========================================================================